Chicago X-Ray Systems, Inc

 
 

Chicago Corix 70 plus - USV - MM -Mobile Unit

Simply select the desired tooth and patient size, and you're on your way.  A powerful microprocessor  automatically chooses the exposure time, according to the type and position of the tooth, patient size, film or digital sensor; with automatic compensation for the main voltage variation.  Manual override is optional if required.
 

Handheld Digital Exposure Timer
A new generation of small-size, 3-digit remote controller with a 1/100 of a second resolution, fully compatible with the fastest film or a film less digital imaging system. The hand held features nine factory preset memories, patient size, manual override, and automatic compensation for main voltage variations.  Digital capability is easily accessible through the click of a switch

The Corix 70 Plus-USV allows the option of configuring a Remote mounted Timer just outside your operatory.

Long Reach Double Positioning Arm
The long reach scissors arm moves effortlessly through a wide area, yet folds
compactly when not in use.  The scissors arm allows ease of positioning and drift free stability.

Corix 70 Plus-USV
Conditions in your practice will dictate where to perform x-ray examinations. With the Corix 70 Plus-USV you can choose the option that best fits your needs

Mobile Stand--mobile unit that allows reaching separate operatory stations on a sturdy and stable rolling stand.
 

Technical Features:

  • Rated Line Voltage: 120V. +/- 10% (109-132VAC), 50/60Hz
  • Power Consumption: 1,050KW max @ 132VAC
  • Rated Tube Potential: 70kVp +/- 7% (Single phase, self rectifying).
  • Rated Tube Current: 8mA +/- 15% @ 120VAC
  • Exposure Time Range: 0.03 to 3.00sec., in steps of 0.01sec.  
       Plus nine preset exposure times, with automatic main voltage compensation.   
       A pre-heating time of 0.15sec., for loading the x-ray tube, is not shown on
       the display.
  • Minimum Focal Spot to Skin Distance: 200mm (8")
  • Radiation Leakage @ 1 m. Less than 50mR/h
       (Technical Factors: 70kVp, 8mA, 1s.).Focal Spot: 0.8 x 0.8mm
  • Total Filtration: 2mm. Al eq.
  • Net Weight: 88lbs
  • Shipping Weight: 102lbs
  • Shipping Dimensions: 2 Boxes; 111 x 34 x 19cm,  86 x 34 x 33cm
  •  

    Item # Description Price
    CXMBL610-08 Corix 70 Plus IntraOral X-Ray Mobile $2,933.82

     


    IntraOs 70 Intraoral X-Ray Mobile

    IntraOs 70 Mobile Intraoral X-Ray is a traditional dental X-ray designed to face the most demanding needs of the dental profession both when using traditional films and digital imaging receptors. The main feature of the system is the great energy offering obtained by combining the voltage potential of 70kVp and the anodic current of 7mA with a wide range of exposure times, from 60ms to 3.2sec.

    The small focal spot, the beam limiting device, and the penetration power of a 70kVp radiation beam, make the best combination of technical parameters which grant sharp images with good radiographic contrast for optimal detail perception.

    Features such as grid control action, and compensation of mains fluctuations, assure consistent film blackening in a wide range of operating conditions, thus granting high diagnostic quality shots.

    This system comes equipped with the AutoSet  timer, offering the best flexibility of use. The AutoSet timer has a flat keyboard and is controlled  by a microprocessor. It features an anatomical selection of technique factors, programmed selections according to tooth type and patient size, manual override (plus, minus) if required, digital display of selected values, dose reduction for digital receptors, pre-settings of film sensitivity, automatic compensation of line voltage fluctuations, full protection against tube over-heating with accurate cool-down control, and X-ray cut-off with "dead-man" control switch functionality.

    Mechanical System:

    The mobile version is composed of:

    • Mobile base: 29" width, 42" Height 24 1/2" depth.
    • Folding arm

    Tube Head

    AutoSet Timer

    • Type: Grid-Control Action
    • Line Voltage: 120 VAC, 50/60Hz
    • Line Current: 8 A
    • High Voltage Potential: 70kVp
    • Anodic Current: 7 mA
    • Focal Spot: 0.8 IEC 336/1982
    • Duty Cycle: 1/30
    • Skin Distance: 20cm
    • Circular Radiation Field: diameter 6 cm
    • Coaxial connection, free rotation in horizontal plane
    • Weight: 14lbs

     

     

     

    • Line voltage: 115VAC, 50/60Hz
    • Microprocessor controlled functionality.
    • Zero-crossing power switching
    • Automatic setting of exposure time by object selection
    • Manual setting of exposure time from 60 ms to 3.2s (plus or minus, 18 steps)
    • Child or Adult patient selection
    • "Low dose" setting for digital receptors
    • Hand-switch provided with 10" coiled cord
    • Automatic compensation of mains fluctuations selectable on request
    • Back-up Timer
    • Timer and hand-switch can be remotely mounted